Tag: contact lens fitting

Not a Steep Learning Curve

[us_post_title]

Using OC-SAG Data to Assess the Suitability For Soft Lens Wear

[us_post_date]

By Gerard Smits

Introduction

This lens fit appears like a complex case, with at least three different factors playing a role. But by ‘simply’ applying a basic set of rules, this case report shows what a difference an Eye Care Professional (ECP) can make in people’s lives. First, by assessing whether an eye is ‘standard’ or not, the ECP can determine whether one of our standard off-the-shelf lenses would be the best lens option, with many benefits including material choice, pricing and – especially – frequency of replacement. ‘Simply’ because the ocular surface is steep and outside the range of the normal eye, choosing a ‘steep lens’ based on sagittal height can make the difference. In addition, there is the challenge (or not) of the lens prescription and material considerations. Not that complicated at all, and, in fact, without a steep learning curve.

Subjective

Tessa came to our practice right in the heart of Amsterdam in the Netherlands. Her prescription is out-of-standard. The feasibility of a standard monthly lens is explored but was questioned. With her high hyperopic prescription, the cosmetic and visual effects of contact lenses over glasses should not be ignored (large visible eyes, aberrations, magnification and sub-optimal visual acuity with glasses).

In the past, she had tried different lens types, with a long track record including, but not limited to Durasoft 3 (Wesley Jessen), Unilab 42C toric (Taiwan) and StabMax (MicroLens). Reproducibility used to be a major problem in the past with custom-made lenses; a problem that has been mostly relegated with current manufacturing techniques. The patient history did not reveal any special or specific considerations or limitations for fitting with a soft lens, other than that she reported having mild dry eyes.

Objective

General slit lamp examination revealed no abnormalities. Refraction of the patient (in this case report we focus on her right eye) came down to: S+4.75=C -5.00 x 90°. Her sim-K values are 7.36/7.14, with eccentricities reported in four different quadrants as: 0.45 superior, 0.71 nasal, 0.58 inferior and 0.35 temporal.

In our contact lens practice, special emphasis is on the shape of the anterior ocular surface. We use a Medmont corneal topographer with a special tool: the E-H software*. Based on the overall corneal topography using that instrument, the peripheral corneal angle measured with the corneal topographer and using the sagittal height calculator – the overall sagittal height of the ocular surface (OC-SAG) over 15mm can be estimated, typically the horizontal meridian values are used.

Assessment

In this case, the E-H 0-180° value was 4097µm. This is well beyond the OC-SAG of the average eye (which is estimated to be in the 3750µm range). In fact, it falls outside the one-standard deviation range (between 3550-3950µm). The fact that this eye falls on the right side of the normal distribution bell curve and even outside the 1-SD range (accountable for 66.7% of all eyes) – indicates that this is an out-of-standard steep eye. Based on keratometry values alone (7.36/7.14), and even in combination with the eccentricity values this time (0.53, which is standard), it is not easy to judge that in the same way. Her corneal diameter (white-to-white) measured with the Medmont was 11.2mm which is slightly smaller than normal.

In addition, the prescription’s astigmatism does not match the corneal astigmatism data (the difference in central keratometry values is only 0.22mm, which represents roughly an astigmatism of 1 diopter, while the refraction indicates 5 diopters of astigmatism). The refractive (rest) cylinder must therefore come from the crystalline lens or another internal astigmatism in the eye.

[us_separator size=”small”][us_image image=”10692″ align=”center”]

Plan

For this patient, it was decided not to start with standard lenses but go directly to an out-of-standard lens – to satisfy both the refractive and ocular surface shape peculiarities of this patient’s eye.

With any toric lens, but certainly, with an astigmatism of this magnitude, good stabilization of the lens on-eye is crucial. Fitting a toric lens is not a disadvantage, in this case, though – but rather an advantage. Since most soft toric lenses have their toricity based on the inside of the lens – this can help stabilization with higher cylindrical values. In other words: a 5D cylinder stabilizes better most probably than a 0.75D cylinder.

In addition, ECPs can choose to fit a slightly steeper lens, to aid stability of the lens on-eye and/or to choose a larger diameter. But for this, it is important to know how steep the eye is. This cannot be judged by central keratometry values and/or eccentricity alone, as this case shows. The eye is exceptionally steep, and to create stability an even steeper lens would be needed. A 7.70 lens was selected in the end, with a lens diameter that is not larger than normal (because of her corneal diameter). But the result is a lens with an out-of-standard high sagittal height value. Using sagittal height values, in this case, makes more sense than using base curve values. 

A Gentle 80 Toric (Mark’Ennovy, Madrid ES) was used in this case: BC 7.70 S+5.00=C -5.00 x 95° Æ14.0. This is an 80% hydrogel lathe-cut soft lens for monthly replacement. The lens with these parameters is calculated to have a CL-SAG of 4100µm. This material, a non-silicone hydrogel (SH) material was preferred over the silicone hydrogel lathe-cut soft lens variant (Saphir Rx ) for monthly replacement, because of wettability issues with the SH lens material.

In Summary

It can be determined that there are 3 categories of soft lenses:

(1) off-the-rack standard soft lenses; 

(2) out-of-standard lenses with extended parameters;

(3) custom-made individually designed lenses. 

For patients like this, the number 1 category is out of range.

First: the refraction is out of the standard lens’ parameter range. In these cases, category 2 lenses can be a huge asset and can prevent drop-out of lens wear. In this case, it was not just a special optical case, but also an ocular surface shape outlier: an exceptionally ‘steep’ eye that desired a special design with a higher-than-normal CL-SAG. Both optically, and in terms of lens comfort, this lens performs better than previous lenses. As a third variable, the silicone hydrogel lenses originally selected did not provide the desired wettability. Switching to a non-SH solved this problem. Careful checks for hypoxic signs on the eye are monitored. But having the option to switch material, whilst staying in the same lens design – can be a huge plus for ECPs when fitting soft specialty lenses.

This case shows that, albeit ‘complex’ at first, applying simple rules towards (1) eye shape, (2) prescription and (3) material properties can make a big difference in contact lens practice.

*the E-H tool in the Medmont topographer can be added simply, using the manufacturers guidelines.

This case study is part of a series focusing on speciality soft lens fitting. Stay tuned for the next installment coming soon. Special thanks to Eef van der Worp for bringing this series together.

[us_separator show_line=”1″ size=”small”]
[us_image image=”10495″ size=”us_600_600_crop” has_ratio=”1″ ratio=”3×4″]

Gerard comes from a long line of opticians (grandparents, father and brother are all opticians). He graduated from the Christiaan Huygenschool (optics) in Rotterdam, and started working as a contact lens specialist in 1985. Since then he has been a member of the ANVC (Dutch Contact Lens Association). He has specialized in both rigid and soft lens fitting and has worked as a consultant for many industry partners, and tested different (specialty) lenses in his practice for the industry. He works as Ralph Optiek in the center of Amsterdam. His experience includes specialty lenses, such as orthokeratology, rigid corneal lenses after refractive surgery and fitting of custom soft lenses. Customization of lens fiting using corneal topography data and multifocal (rigid corneal and soft) lens fitting are two of the main challenges, but also opportunities he sees for your profession.

[us_separator size=”small” show_line=”1″]
[us_post_navigation in_same_term=”1″]
[us_sharing providers=”facebook,twitter,linkedin” align=”right”]
Tags: , , , , , ,

Avoiding a Corneal Transplant in a Young Patient

[us_post_title][us_post_date]

By William Skoog OD, FAAO

Affiliations: Illinois College of Optometry

Background

Corneal opacity is the fifth leading cause of blindness worldwide¹. A common cause of corneal opacity is infectious keratitis. In developed countries, contact lenses are the leading cause of infectious keratitis while in developing countries it is from corneal trauma during agriculture work¹. The most common organism responsible for microbial keratitis in contact lens wear is Pseudomonas aeruginosa². When the visual axis is involved, a cornea transplant may be required to improve vision. Corneal transplants have a higher rate of rejection in younger patients³. 

Case Description

A 19-year-old Hispanic female presented for a gas permeable contact lens fitting. Her ocular history was remarkable for microbial keratitis OS caused by Pseudomonas following extended wear of an unknown monthly soft contact lens. 

Entering visual acuities were 20/20 OD with correction and 20/200 OS without correction which improved to 20/40 with pinhole. All other entrance testing was unremarkable.

Slit lamp examination revealed a 6 mm round, central stromal scar with significant stromal thinning OS (Figure 1). 

Simulated keratometry measurements were taken from tomography and were 40.75/43.0 @ 89 OD (Figure 2), 33.75/35.50 @ 99 OS (Figure 3). 

Manifest Refraction was -7.00-1.50×170 OD with a distance acuity of 20/20 and +2.00 sphere OS with a distance acuity of 20/80. 

Due to the decrease in VA secondary to a central corneal opacity a diagnostic scleral lens was applied to the left eye to determine visual potential. The initial diagnostic lens that was applied to OS was a prolate lens (Table 1A) that showed excessive central fluid reservoir clearance (Figure 4A). Her VA improved to 20/30 with an over-refraction. Since there was a vast improvement in VA with a scleral lens, the decision was made to peruse wearing a scleral lens OS. The final lens order was changed to an oblate design (Table 1B) which improved the central fluid reservoir clearance (Figure 4B). 

Optimum Extreme, a high-DK material was selected to allow the compromised cornea to receive as much oxygen as possible while she was wearing the lens. Since the initial corneal insult happened because of non-compliance, the importance of contact lens hygiene, wearing schedule, cleaning regimen and proper filling solution was discussed at length with her and her parents. 

She was closely followed over the first three months of scleral lens wear to ensure appropriate wear.

[us_separator size=”small”][us_image image=”10663″ size=”us_600_0″ align=”center”]

Figure 1: Slit lamp photo showing a 6mm round central stromal scar OS secondary to microbial keratitis.

[us_separator]
[us_image image=”10665″ size=”us_600_0″ align=”center”]

Figure 2: Axial map showing with-the-rule astigmatism in the right eye.

[us_image image=”10664″ size=”us_600_0″ align=”center”]

Figure 3: Axial map showing central corneal flattening secondary to microbial keratitis in the left eye.

[us_separator]
[us_image image=”10666″ size=”us_600_0″ align=”center”]

Figure 4A: An AS-OCT of a prolate diagnostic scleral lens OS showing excessive central vault with adequate mid-peripheral vault.

[us_image image=”10667″ align=”center” size=”us_600_0″]

Figure 4B: An AS-OCT of an oblate scleral lens design OS showing adequate central clearance after ~6 hours of wear.

[us_separator]
Type Power BC Diameter Edge Design
Onefit 2.0 Plano 8.4 14.9 std Prolate

Table 1A. Initial diagnostic scleral lens OS that showed excessive central clearance

[us_separator]
Type Power BC Diameter Edge Design Material
Onefit 2.0 +6.50 8.4 14.9 std Oblate – CCR 230 Optimum Extreme

Table 1B. Final scleral lens design OS 

[us_separator size=”small”]

Conclusions

Corneal scarring after microbial keratitis leads to vision loss secondary to corneal opacity and irregular astigmatism⁴. The two most common options to improve vision in these cases are corneal transplantation or a specialty contact lens. Corneal gas permeable or scleral contact lenses should be considered if they are able to provide the patient with functional vision as was demonstrated in this case.

[us_separator show_line=”1″ size=”small”]
1. Ting, D.S.J., Ho, C.S., Deshmukh, R. et al. Infectious keratitis: an update on epidemiology, causative microorganisms, risk factors, and antimicrobial resistance. Eye 35, 1084–1101 (2021). https://doi.org/10.1038/s41433-020-01339-3
2. Tan, S., Walkden, A., Au, L. et al. Twelve-year analysis of microbial keratitis trends at a UK tertiary hospital. Eye 31, 1229–1236 (2017). https://doi.org/10.1038/eye.2017.55
3. Gulias-Cañizo R, Gonzalez-Salinas R, Hernandez-Zimbron LF, Hernandez-Quintela E, Sanchez-Huerta V. Indications and outcomes of pediatric keratoplasty in a tertiary eye care center: A retrospective review. Medicine (Baltimore). 2017 Nov;96(45):e8587. doi: 10.1097/MD.0000000000008587. PMID: 29137083; PMCID: PMC5690776.
4. Singh K, Jain D, Teli K. Rehabilitation of vision disabling corneal opacities: is there hope without corneal transplant? Cont Lens Anterior Eye. 2013 Apr;36(2):74-9. doi: 10.1016/j.clae.2012.10.085. Epub 2012 Nov 27. PMID: 23199902.
[us_separator show_line=”1″ size=”small”]
[us_image image=”10662″ size=”us_600_600_crop” has_ratio=”1″]

Dr. Skoog is a clinical instructor at the Illinois College of Optometry in the cornea and contact lens department, where he previously completed a residency. He lectures on anterior segment disease and contact lenses. He has no disclosures.

[us_separator size=”large”]
[us_separator size=”small” show_line=”1″]
[us_post_navigation in_same_term=”1″]
[us_sharing providers=”facebook,twitter,linkedin” align=”right”]
Tags: , , , , , , , ,

Using a Customized Soft Lens

[us_post_title]

Sometimes ‘flat out’ the best option for a flat eye

[us_post_date]

By Gerard Smits

Introduction

The vast majority of the lenses we fit in contact lens practice are soft lenses. The amount of attention that goes into ‘fitting’ soft lenses is limited in contrast to the focus on scleral lens fitting, corneal lens fitting, and orthokeratology, for instance. This case shows how taking control of the fitting process can sometimes enhance someone’s life and upgrade one’s vision in relation to quality of life. 

Subjective

Mary came to our practice right in the heart of Amsterdam in the Netherlands. Her wish was to wear lenses in addition to the glasses that she used, with a preference for daily disposables if possible. The anamneses did not reveal any specific considerations or limitations for fitting with a soft lens. Patient complains of moderate dry eye symptoms.

Objective

General slit lamp examination revealed no abnormalities. We focus on the right eye of the patient for this case report. Refraction of the right eye came down to S+2.75, with a corneal diameter (white-to-white with the Medmont topographer) of 12.0mm with an overall average eccentricity of over four quadrants of 0.43.

In our contact lens practice, emphasis is on the shape of the entire anterior ocular surface, not just the central cornea. We use a Medmont corneal topographer for this reason, but with a special tool: the E-H software¹. 

Based on the overall corneal topography using that instrument, combined with the peripheral corneal angle measured with the topographer and using the sagittal height calculator – the overall sagittal height of the ocular surface (OC-SAG) over 15mm can be estimated, typically the horizontal meridian values are used.

In this case, the E-H 0-180° value 3524 microns. This figure is well below the OC-SAG of the average eye, which is estimated to be in the 3750 micron range. This eye also falls outside the one-standard deviation range (1-SD), which is between 3550-3950 microns.  

[us_separator][us_image image=”10467″ size=”us_768_0″ align=”center”][us_separator]

Assessment

The fact that this eye falls outside (on the left side) of the normal distribution bell-curve and even outside the 1-SD range – indicates that this is an exceptionally flat eye. Based on keratometry values alone (8.09/7.88) it is not easy to make the same judgement, especially as the eye has an average e-value and standard corneal diameter. In cases like this, measuring the full ocular surface over at least the diameter of the soft lens (or the benchmark 15mm is used as a standard often) can be very insightful. 

For this patient, several standard soft lenses were tried, including Dailies Aqua Comfort Plus (Alcon) BC 8.70 Ø 14.0, Clariti 1 Day (CooperVision) BC 8.60 Ø 14.1, Dailies All Day Comfort BC 8.60 Ø 13.80 and monthly disposable Biofinity (Alcon) BC8.60 Ø 14.0 and Fresh Excellent (Microlens/Wölk) BC 8.3 Ø 13.6. All lenses failed because of lack of movement and mobility, in addition to poor comfort.

Plan

Patient ultimately was refitted with a custom soft lens based on the patient’s corneal topography. The Medmont file was­­­ uploaded to the manufacturer and an individual monthly disposable silicone hydrogel lens was created for this patient. The Indivisual (Menicon, Emmen NL) was manufactured in the Definitive 74 material (Efrofilcon A V3) with a sagittal height of the lens (CL-SAG) of 2900 micron and a tangential peripheral shape (tangent angle 54°) S+2.50 in a Ø 14.1. 

The lens design is similar to how many scleral lens designs are configured these days. The lens fit procedure would also be close to fitting a scleral lens: changing the tangent angle would ‘lift’ the lens, decreasing the lens to a lower sagittal height. For this patient, this lens design resulted in an optimal lens fit with excellent movement – which is believed to be the result of the tangential peripheral shape of the lens in conjunction with the customized CL-SAG height. 

In Summary

Using the overall sagittal height of the ocular surface over a 15mm chord can give the eye care practitioner (ECP) valuable information about when to start with a flat or steep lens (a low or high CL-SAG) using the Pacific University SAG-charts, or when to divert to an out-of standard or a customized lens – like in this case. This will save time, on behalf of the patient and the ECP, from fewer trial lenses used and increased overall satisfaction. Consequently, this can potentially prevent drop-out of soft lens wear. 

The initial lens type requested by the patient, daily disposables, come in a variety of brands but are limited in size and parameters. It is the task of the ECP to assess and decide which lens is suitable for a given eye and which is not. This case perfectly illustrates the decision-making process, and a custom-made lens in a silicone hydrogel material with a monthly replacement schedule is the ideal solution for the patient. The patient still desires a daily disposable lens, if possible, but that is not feasible in this case. As ECPs we must be clear about this. In regard to the patient’s complaint of moderate dry eye symptoms – as ECPs we must be careful with that. Any form and type of eye discomfort can be easily mistaken for dry eye. But often, lens fit pro­­blems are partially or fully the root cause of discomfort issues, and lens fit issues need to be addressed first. The ‘take-home message’ in this case was pretty obvious: a flat lens for this flat eye was the best technical option for this patient.  

¹the E-H tool in the Medmont topographer can be added simply, using the manufacturer’s guidelines. 

This case study is part of a series focusing on speciality soft lens fitting. Stay tuned for the next installment coming soon. Special thanks to Eef van der Worp for bringing this series together.

[us_separator show_line=”1″ size=”small”]
[us_image image=”10495″ size=”us_600_0″ has_ratio=”1″ ratio=”3×4″]
[us_separator size=”small”]

Gerard comes from a long line of opticians (grandparents, father and brother are all opticians). He graduated from the Christiaan Huygenschool (optics) in Rotterdam, and started working as a contact lens specialist in 1985. Since then he has been a member of the ANVC (Dutch Contact Lens Association). He has specialized in both rigid and soft lens fitting and has worked as a consultant for many industry partners, and tested different (specialty) lenses in his practice for the industry. He works as Ralph Optiek in the center of Amsterdam. His experience includes specialty lenses, such as orthokeratology, rigid corneal lenses after refractive surgery and fitting of custom soft lenses. Customization of lens fiting using corneal topography data and multifocal (rigid corneal and soft) lens fitting are two of the main challenges, but also opportunities he sees for your profession.

[us_separator size=”small” show_line=”1″]
[us_post_navigation in_same_term=”1″]
[us_sharing providers=”facebook,twitter,linkedin” align=”right”]
Tags: , , , , , , ,

The Changing Face of Contact Lens Fitting

[us_post_title][us_post_date][us_image image=”10284″ size=”us_600_0″ align=”center”]

By Martin Conway

In the October 2022 edition of Contact Lens Spectrum, Ed Bennett’s report on the GP and Custom Soft Annual Report 2022 provided a comprehensive insight into the prescribing habits of top US practitioners. Scanning the list of subscribers, they are mainly Key Opinion Leaders and researchers, so this may not necessarily be indicative of what is going on in the rest of the US market, much less in diverse markets around the world. It does, however, give a good indication of the direction the contact lens industry will take over the next ten years or so.

We also see the growth of interest in the control of myopia with both soft and GP ortho-k, as well as the continued use of scleral lenses – none of which should surprise us.

Another prominent trend is the increased use of complex optics such as wavefront-guided lens designs and the correction of Higher Order Aberrations, requiring complex calculations to achieve success. All of which need accurate digital data in order to be effective.

Practitioners fitting ortho-k lenses typically do so with the aid of a topographer and not a fitting set. The recent increased use of scleral lenses has been facilitated, in great part, by the routine use of OCT during normal optometric practice. So we see that increased technology use has already occurred over the past decade or two.

Specialist lens practitioners have long since exchanged their keratometers for topographers; the data produced is then converted by laboratories’ algorithms to provide a points file which is fed directly into a lathe. This delivers an almost guaranteed first-time fit and has replaced the need for fitting sets for ortho-k and regular corneal lenses. Laboratories also benefit by not having to provide dozens of fitting sets, many of which are seldom used, and the practitioner benefits from fewer refits and reduced chair time.

The information provided by a Placido Ring-based instrument does not cover the entire cornea, and although a vast improvement on the keratometer, a certain amount of extrapolation is required in order to give information up to and beyond the limbus.

Enter the Corneal Profilometer. For some years, the suppliers of these instruments have been extolling the virtue of their advantages over conventional topography. These have had limited success in Europe, but the US market appears to have accepted the concept more readily than elsewhere. This is perhaps because of the more rapid uptake of scleral lenses in that market, and it would appear from the report that their use is now becoming more mainstream.

The information produced by these new instruments allows for more personalised designs to be created, incorporating edge profiles that exactly match the patient. Again, the information collected by the instrument provides the raw data for an algorithm that ultimately feeds the lathe with data in order to produce the finished lens.

Scheimpflug-based instruments are also available on the market, which will, not only collect data over the entire cornea and provide 360-degree imagery but also download that data to an online lens design module that, in turn, instructs the lathe directly.

There is now a small but growing interest in fitting intra-limbal designs for regular corneae because of the additional comfort they provide for the new wearer. Experience with ortho-k designs, which tend to be slightly larger and more comfortable than normal GP designs, has persuaded practitioners that larger diameter lenses might offer greater comfort than conventional designs for their regular patients.

Although these designs have been suggested for some time, practitioners found the unfamiliar fluorescein pattern produced by these larger lenses confusing, and apart from a few dedicated followers, their use has been limited.

With accurate data from the limbal area and beyond, this modality should develop further as algorithms take over the need to learn new skills. Those practitioners already using this type of lens describe the comfort as similar to soft lenses within two hours of wear, yet with the visual acuity typically associated with GP lenses.

Looking forward, I can see that practices wishing to develop their speciality contact lens business will need to commit to investing in instrumentation within their practice. Either in the new generation of topographers, which can provide a composite picture of the complete cornea and beyond, one of the Corneal Profilers, or Scheimpflug anterior chamber imaging systems.

As lens designs develop and provide increased comfort and vision for myopia management, ectasia, and other ametropias, more opportunities will open up for the profession. These opportunities will require accurate data to be provided to the laboratory, and consequently, laboratories will need to be set up to receive this digital data and apply it to their new and existing designs.

Practices that do decide to invest in their speciality lens business can look forward to reaping the benefits that these instruments permit. They will differentiate themselves as patients appreciate the level of investment and enjoy the increased comfort and visual benefits they provide.

These developments will not come about overnight. Ed Bennett’s report gives a snapshot of what is happening within some very sophisticated and specialised practices, and the US market has a significant influence on how the worldwide market will develop.

We can look forward to exciting developments over the next few years, led by diagnostic technologies that will change the way bespoke contact lenses of all types are supplied.

[us_separator show_line=”1″ size=”small”]
[us_image image=”8848″]
[us_separator size=”small”]

Martin Conway has over 40 years’ experience in the contact lens field as a qualified Contact Lens Optician. He is registered with the UK General Optical Council on the Speciality Contact Lens Register. Martin is a fellow of the British Contact Lens Association (FBCLA), and The International Association of Contact Lens Educators (FIACLE). He has served in the Professional Services role as an educator and clinical adviser on behalf of both Sauflon and CIBA, and now acts as Professional Services Consultant for Contamac. Martin has lectured extensively in Europe, Asia, Russia, North and South America and the Middle East.